MULTIDOS
RS232 Interface
Application Afterloading
22
D177.131.2/0
11.3.3 Read maximum value
The response to the DMc telegram is
DMc•m.mmE±ee
where c is the channel number. The transmitted value is the maximum measurable value of the doserate or
current measurement. The value is formatted in scientific notation.
11.3.4 Read Resolution
The response to the DRc telegram is
DRc•0.mmmE±ee
where c is the channel number. The transmitted value is the resolution of the actual measuring mode for the
channel. The number of digits following the decimal point may vary from 1 to 3.
11.3.5 Measuring unit
Telegram
Response
DU
DU
u
Reads the absolute unit of measurement of the active measuring mode. Possible values of
u
:
when MULTIDOS is set to radiological units:
Gy Gy/s Gy/min Gy/h
when MULTIDOS is set to electrical units:
C A
11.4 Calibration Factors
Telegram
Description
Response
CR
s
A
aa
Set/Read the status of the channels of set
s
.
aa
is a 2 digit
decimal number, whose set bits, when converted to binary
representation, signal the following status:
.0
Þ
channel 1 of the rectum probe is active
.1
Þ
channel 2 of the rectum probe is active
.2
Þ
channel 3 of the rectum probe is active
.3
Þ
channel 4 of the rectum probe is active
.4
Þ
channel 5 of the rectum probe is active
.5
Þ
channel of the bladder probe is active
.0 is the least significant bit
aa
can take values from 00 to 63. from 00 to 63.
CRsAaa
CR
s
F
cf
Set/Read the calibration factor
f
of channel
c
of set
s
CRsFccf
CR
s
N
name
Set/Read the name of set
s
CRsNname
CR
s
T
dtyp
Set/Read the ID of detector
d
of set
s
.
typ
: string with max.15 characters (ASCII; 20H to 7FH)
Space characters are allowed.
Ä, Ö, Ü, ä, ö, ü, ß, µ are supported.
d
: R (rectum probe), B (bladder probe)
CRsTcctyp
CR
s
D
dd.mm.yyyy
Set/Read the calibration date of set
s
CRsDdd.mm.yyyy
CR
s
U
x
Update (
x
= 1)/Read (no
x
) the result of the check sum pro-
tection of set
s
CRsUx
CR
s
B
bbbbb
Set/Read the block check value
bbbbb
of set
s
CRsBbbbbb