Remote Control Commands
R&S
®
SMC100A
284
Operating Manual 1411.4060.32 ─ 04
Parameters:
<Delay>
float
Range:
10 ns to 100 s
Increment: 20 ns
*RST:
1 ms
Default unit: ns
Example:
PULM:DOUB:DEL 22 us
22 us elapse between the beginning of the first pulse and the
beginning of the second pulse in double-pulse mode.
[:SOURce<hw>]:PULM:DOUBle:STATe
<State>
The command enables/disables double pulse generation. The two pulses are generated
in one pulse period.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
PULM:DOUB:STAT ON
double-pulse mode is enabled.
[:SOURce<hw>]:PULM:DOUBle:WIDTh
<Width>
The command sets the width of the second pulse in case of double pulse generation.
Parameters:
<Width>
float
Range:
5ns to 100s
Increment: 20ns
*RST:
1ms
Default unit: ns
Example:
PULM:DOUB:WIDT 33 us
sets a width of 33 us for the second pulse.
[:SOURce<hw>]:PULM:MODE
<Mode>
Sets the mode of the pulse generator.
Parameters:
<Mode>
SINGle | DOUBle
SINGle
Enables single pulse generation.
DOUBle
Enables double pulse generation. The two pulses are generated
in one pulse period.
*RST:
SINGle
SOURce Subsystem