Data Output Format with 22 Characters
When data is output with an ID code, the 6-character code precedes the 16-character string
described above. The code identifies the subsequent value.
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22
I
I
I
I
I
I
+
*
D
D
D
D
D
D
D
D
*
U
U
U
CR LF
*
*
*
*
*
-
.
.
.
.
.
.
.
.
*
*
*
*
*
*
*
*
*
*
*
*
0
0
0
0
0
0
I:
ID code character
1)
U:
Unit symbol
*:
Space
CR:
Carriage return
D:
Digit or letter
LF:
Line feed
Special Codes
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22
S
t
a
t
*
*
*
*
*
*
*
*
-
-
*
*
*
*
*
*
CR LF
H
*
L
*
*:
Space
H:
Overload
- -:
Final readout mode
L:
Underload
Error Codes
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22
S
t
a
t
*
*
*
*
*
E
R
R
*
#
#
#
*
*
*
*
CR LF
*:
Space
# # #:
Error code number
Characters for ID
code I
Meaning
Stat
Status
G
Gross
G#
Gross G calculated
T
Tare T
T1
Tare T1
N
Net N
N1
Net N1
Data Input Format
The computer connected to the inter-
face can be used to send commands to
control scale functions and application
programs. These are control commands
and may have different formats. Control
commands consist of up to 4 characters.
Each character must be transmitted
according to the settings configured in
the operating menu for data transmis-
sion.
Format for Control Commands
Format 1:
Esc
!
CR
LF
Esc:
Escape
CR:
Carriage return (optional)
!:
Command character
LF:
Line feed (optional)
Command
character ! Meaning
K
Weighing mode 1
L
Weighing mode 2
M
Weighing mode 3
N
Weighing mode 4
O
Block keys
P
R
Unblock keys
S
Restart
U
Tare the scale (tare only)
V
Zero the scale
W
External calibration
10