4-3. Synchronous GATE signal data acquisition commands
Type of command
GSTRT
Start to store gate synchronous data. (Just after turn on power,
it's stop condition)
Entering this command after preparation of section 4-2,
CT16-ER2 starts data acquisition synchronized gate signal.
It stores signal data from current data address to end data address
synchronized by Gate signal.
Inputting "STOP" command, data acquisition stops immediately.
Type of command
GESTRT
Start to store gate edge synchronous data. (Just after turn on power,
it's stop condition)
Entering this command after preparation of section 4-2,
counter gate opens at the first falling edge of gate signal,
CT16-ER2 starts synchronous falling edge of gate signal data acquisition.
It stores signal data from current data address to end data address
synchronized by falling edge of Gate signal.
Inputting "STOP" command, data acquisition stops immediately.
4-4. Synchronous internal clock data acquisition commands
In this mode, data acquisition is synchronized by clock generated by CT16-ER2.
It's necessary to preset the value of clock ON time and clock OFF.
External gate input signal works gate signal during data acquisition operation.
If you're going to store data synchronized clock gate signal only, please set gate signal to ON.
(GATE ON = input OPEN:positive /Default setting is positive.)
4-4-1. Set and read synchronous Timer RUN time
Type of command
GTRUNddd・・・
Set synchronous timer data RUN time
(1μs~4,294,967,295μs : μs units.)
To make sure of time enough to store data, it's recommended that
the sum of gate RUN time and STOP is over 10,000(=10ms).
GTRUN?
Read synchronous timer data RUN time
Reply
20000
Read by μs units. Ex. 20,000 = 20ms
4-4-2. Set and read synchronous timer OFF time
Type of command
GTOFFddd・・・
Set timer synchronous data STOP time
(1μs~4,294,967,295μs : μs units. If you input 0, set about 200ns.)
To make sure of time enough to store data, it's recommended that
the sum of gate RUN time and STOP is over 10,000(10ms).
Type of command
GTOFF?
Read the timer synchronous data STOP time
Reply
20000
Read μs units. Ex. 20,000 = 20ms
27/43