Programmer’s Guide
5-23
Using Status Registers
The Analyzer's Status Register Sets
Measuring Status Register Set
The Measuring Status register set monitors conditions in the analyzer's
measurement process.
Bits in the Measuring Status condition register are set to 1 under the
following conditions:
Channel 1
Measuring
(bit 0) is set to 1 while the analyzer is collecting
measurement data on channel 1.
Channel 2
Measuring
(bit 1) is set to 1 while the analyzer is collecting
measurement data on channel 2.
Averaging Status Register Set
The Averaging Status register set monitors conditions in the analyzer's
measurement process when the trace averaging function is in use.
Bits in the Averaging Status condition register are set to 1 under the
following conditions:
Measurement
Channel 1
Averaging
(bit 0) is set to 1 while the analyzer is sweeping on
measurement channel 1 and the number of sweeps
completed (since “average restart”) is less than the
averaging factor.
Measurement
Channel 2
Averaging
(bit 1) is set to 1 while the analyzer is sweeping on
measurement channel 2 and the number of sweeps
completed (since “average restart”) is less than the
averaging factor.
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...