3 SCPI
3.3 Instrument-specific Commands
42
[ User TCold ]
:SENSe:CORRection:TCOLd:USER:VALue
Function description:
This command is used to set/query the user cold temperature value.
Setting format:
:SENSe:CORRection:TCOLd:USER:VALue <temperature>
Query format:
:SENSe:CORRection:TCOLd:USER:VALue?
Parameter description:
The unit is Kelvin (K), Celsius (CEL), or Fahrenheit (FAR). This value is used
when the user cold temperature switch is enabled; this value is not used
when it is read out from a smart noise source. The default unit returned by the
query is K.
Range: [0, 29650000.00] K
Example:
:SENSe:CORRection:TCOLd:USER:VALue 290K
:SENSe:CORRection:TCOLd:USER:VALue?
Key path:
Front panel
【
ENR
】
—>[TCold]—>[User TCold]
Reset status:
296.5K
3.3.4 Loss Compensation
The loss compensation corresponds to the
【
Loss Comp
】
key on the front panel. It is used to set the
related loss compensation parameters before and after the device under test during measurement. A
single fixed loss value that is applied to all frequencies can be specified, or the loss compensation table
applying to the frequencies within the entire frequency range/corresponding to the loss can be set. The
commands include loss compensation on-off, type, fixed loss compensation value, loss compensation
table, and temperature compensation value.
3.3.4.1 Settings before DUT
[ Loss Compensation Control before DUT ]
:SENSe:CORRection:LOSS:BEFore[:STATe]
Function description:
This command is used to set/query the loss compensation status before the
DUT.
Setting format:
:SENSe:CORRection:LOSS:BEFore[:STATe] OFF|ON|0|1
Query format:
:SENSe:CORRection:LOSS:BEFore[:STATe]?
Parameter description:
ON|1: Loss compensation on
OFF|0: Loss compensation off
Example:
:SENSe:CORRection:LOSS:BEFore:STATe OFF
:SENSe:CORRection:LOSS:BEFore:STATe?
Key path:
Front panel
【
Loss Comp
】
—>[Before DUT]
Reset status:
Off
[ Loss Compensation Mode before DUT ]
:SENSe:CORRection:LOSS:BEFore:MODE
Function description:
This command is used to set/query the loss compensation mode before the
DUT.
Setting format:
:SENSe:CORRection:LOSS:BEFore:MODE OFF|FIXed|TABLe