3 SCPI
3.3 Instrument-specific Commands
38
Setting format:
:SENSe:CORRection:ENR:CALibration:TABLe:SNS
Parameter description:
When the ENR data is loaded, any measurement in progress will be reset.
This command will give a setting conflict prompt when there is no SNS smart
noise source connection.
Example:
:SENS:CORR:ENR:CAL:TABL:SNS
[ Measurement ENR Table ]
:SENSe:CORRection:ENR[:MEASurement]:TABLe:DATA
Function description:
This command is used to set/query the data in the current measurement ENR
table.
Setting format:
:SENSe:CORRection:ENR[:MEASurement]:TABLe:DATA <freq>,<val>{,<freq>,<val>}
Query format:
:SENSe:CORRection:ENR[:MEASurement]:TABLe:DATA?
Return value:
<freq>,<value>{,<freq>,<value>} The default units are Hz and dB
Parameter description:
<Freq> Frequency, the frequency units that can be input include Hz, kHz, MHz, GHz, and
the default unit is Hz
<value> ENR value in dB
Range: [1, 81] data groups
Example:
:SENSe:CORRection:ENR:MEASurement:TABLe:DATA 1GHz,15.31dB,2G,15.5
:SENSe:CORRection:ENR:MEASurement:TABLe:DATA?
Key path:
Front panel
【
ENR
】
—>[Common Table|Off]—>[Measurement ENR Table]—>[Edit]
[ Number of Measurement ENR Table ]
:SENSe:CORRection:ENR[:MEASurement]:TABLe:COUNt?
Function description:
This command is used to query the number of ENR values in the current
calibration ENR table.
Query format:
:SENSe:CORRection:ENR[:MEASurement]:TABLe:COUNt?
Return value:
Integer
Parameter description:
Range: [0, 81]
Example:
:SENSe:CORRection:ENR:MEASurement:TABLe:COUNt?
[ Measurement ENR Table Type ]
:SENSe:CORRection:ENR[:MEASurement]:TABLe:ID:DATA
Function description
: This command is used to set/query the noise source type in the current
measurement ENR table.
Setting format:
:SENSe:CORRection:ENR[:MEASurement]:TABLe:ID:DATA <ID>
Query format:
:SENSe:CORRection:ENR[:MEASurement]:TABLe:ID:DATA?
Return value:
Noise source type.
Parameter description:
<ID> Maximum 12 characters (such as: 346C)