3 SCPI
3.3 Instrument-specific Commands
85
Parameter description:
OFF|0: Continuous sweep off. At this moment, the instrument keeps a waiting
state and starts to sweep until it receives the command of setting
continuous on. It will automatically return to the waiting state after
completing a single sweep.
ON|1: Continuous sweep on. At this moment, another measurement starts
immediately after the completion of one sweep.
Return value:
Microwave maximum attenuation when the calibration is made
Example:
:INITiate:CONTinuous OFF
:INITiate:CONTinuous?
Reset status:
Continuous sweep
Key path:
Front panel
【
Sweep
】
—>[Sweep Mode]
[ Restart ]
:SENSe:RENEw:SWEep
Function description:
This command is used to renew a sweep.
Setting format:
:SENSe:RENEw:SWEep
Example:
:SENSe:RENEw:SWEep
Key path:
Front panel
【
Sweep
】
—>[Restart]
[ Restart ]
:INITiate[:IMMediate]
Function description:
This command is used to renew a sweep.
Setting format:
:INITiate[:IMMediate]
Example:
:INIT
Key path:
Front panel
【
Sweep
】
—>[Restart]
[ Restart ]
:INITiate:RESTart
Function description:
This command is used to renew a sweep.
Setting format:
:INITiate:RESTart
Example:
:INIT:REST
Key path:
Front panel
【
Sweep
】
—>[Restart]
3.3.12.2 Manual Measurement
[ Manual Measurement Noise Source Control ]
:SENSe:MANual:NOISe[:STATe]
Function description:
This command is used to set/query the on and off states of the noise source.
Setting format:
:SENSe:MANual:NOISe[:STATe] OFF|ON|0|1
Query format: