3 SCPI
3.3 Instrument-specific Commands
44
[ Number of Input Loss Table before DUT]
:SENSe:CORRection:LOSS:BEFore:TABLe:COUNt?
Function description:
This command is used to return the number of input loss compensation before
DUT.
Query format:
:SENSe:CORRection:LOSS:BEFore:TABLe:COUNt?
Parameter description:
Range: [0, 201] dB
Example:
:SENSe:CORRection:LOSS:BEFore:TABLe:COUNt?
[ Loss Compensation Temperature before DUT ]
:SENSe:CORRection:TEMPerature:BEFore
Function description:
This command is used to set/query the temperature of the loss compensation
before DUT.
Setting format:
:SENSe:CORRection:TEMPerature:BEFore <temperature>
Query format:
:SENSe:CORRection:TEMPerature:BEFore?
Parameter description:
The unit is Kelvin (K), Celsius (CEL), Fahrenheit (FAR) or K, C, F. The default
unit returned by the query is K.
Range: [0, 29650000.00] K
Example:
:SENSe:CORRection:TEMPerature:BEFore 1000K
:SENSe:CORRection:TEMPerature:BEFore?
Key path:
Front panel
【
Loss Comp
】
—>[Temp (Before)]
Reset status:
290.0K
3.3.4.2 After DUT Setup
[ Loss Compensation Control after DUT ]
:SENSe:CORRection:LOSS:AFTer[:STATe]
Function description:
This command is used to set/query the loss compensation status after DUT.
Setting format:
:SENSe:CORRection:LOSS: AFTer [:STATe] OFF|ON|0|1
Query format:
:SENSe:CORRection:LOSS: AFTer [:STATe]?
Parameter description:
ON|1: Loss compensation on
OFF:|0: Loss compensation off
Example:
:SENSe:CORRection:LOSS: AFTer:STATe OFF
:SENSe:CORRection:LOSS: AFTer:STATe?
Key path:
Front panel
【
Loss Comp
】
—>[After DUT]
Reset status:
Off
[ Loss Compensation Mode after DUT ]
:SENSe:CORRection:LOSS:AFTer:MODE
Function description:
This command is used to set/query the loss compensation mode after DUT.
Setting format:
:SENSe:CORRection:LOSS:AFTer:MODE OFF|FIXed|TABLe