Filter (Averaging) Group
SENSe:AVERage:STATe <ON|OFF> or <1|0>
Enables or disables the filter (rolling average) facility.
SENSe:AVERage:STATe?
Returns “0” (OFF) or “1” (ON).
SENSe:AVERage:COUNt <
No. of readings>
Sets the number of readings to be filtered or averaged (1 - 32).
SENSe:AVERage:COUNt?
Returns the number of readings to be filtered (averaged).
Selecting Filter on will automatically deselect the Settling facility.
Settling Algorithm Group
SENSe:SETTling:STATe <ON|OFF> or <1|0>
Enables or disables the Settling facility.
SENSe:SETTling:STATe?
Returns “0” (OFF) or “1” (ON).
SENSe:SETTling:COUNt <
No. of readings>
Sets the maximum number of readings to be taken (2 - 999).
SENSe:SETTling:COUNt?
Returns the maximum number of readings selected.
SENSe:SETTling:LIMit <
No. of digits>
Sets the maximum number of display di gits by which two consecutive readings
may differ, for the measurement to be defined as ‘settled’.
SENSe:SETTling:LIMit?
Returns the maximum number of display digits selected.
Selecting the Settling facility will:
Deselect the Filter facility.
Change Average (AVE) Current mode to Positive (+I) Current mode.
Cause an error message if in Fast mode.
Selecting Filter on, Average Current or Fast mode will automatically deselect the
Settling facility.
Open Circuit Voltage Limit Group
SOURce:VOLTage:LIMit:LEVel <OFF|0|20|50>
Sets the open circuit limit to OFF, 20mV or 50mV.
Open Circuit Limits are not available if the 3KOHM or 30KOHM range is selected,
and attempting to set 20mV or 50mV will generate an “execution error” and the
command will be ignored.
Attempting to set 20mV or 50mV with AUTO1 or AUTO2 selected will be
successful, but autorange mode will be turned off and the instrument forced to the
current manual resistance range.
The DO5003 does not have this facility, and attempting to set it will generate an
“execution error” and the command will be ignored.
SOURce:VOLTage:LIMit:LEVel?