™/S, MOLBLOC TERMINAL OPERATION AND MAINTENANCE MANUAL
© 1995 - 2010 Fluke Corporation, DH Instruments Division
Page 134
TARESET(=)
Purpose
To automatically
tare
the upstream and downstream RPT or set the
tare
to a user
defined value. Not available on a S.
Syntax
“TARESET=tare
“TARESET”
Defaults
“TARESET=0”
Arguments
tare:
The
tare value
(-9999 to 9999 Pa).
Remarks
“TARESET” will automatically
tare
the only if the
tare
results are within
the limits given (see the “TARE” command). Use the “TARESETUP” command to
prepare the unit to be
tared
, and the “TARE” command to check if conditions are
valid. After you have completed taring the unit, use the “TARESETUP=OFF”
command or the “ABORT” command to return the unit to normal operating
conditions. You can also manually set the tare value using the “TARESET=tare”
command. All pressure values are always in Pascal [Pa]. While the
tare
cycle is
executing, a ‘b’ (“busy”) will appear in the third character position of the “FR” and the
“SR” command replies.
The normal order of remote commands for a
tare
operation:
“TARESETUP=UP” or “TARESETUP=DN” configures the for a
tare
.
“TARE” queries the for current conditions until it is ready for
tare
.
“TARESET”
tares
the when the “TARE” query has shown it is ready.
“TARESETUP=OFF” returns the to normal measurement mode.
NOTE:
This command typically takes 5 to 7 seconds to complete if performing a
tare
(no arguments given).
Example
Command:
Reply:
Reply:
“TARESET”
“66 Pa”
“66 Pa, 7 Pa”
Errors
ERR# 6:
ERR# 25:
ERR# 52:
An argument is invalid.
The
tare
values
exceed the
tare
limit of 3,000 Pa.
Cannot tare in A+B or A/B mode.
See Also
3.4.4.1 , “TARE”, “TARESETUP”, “ABORT”
TCAL(=)
Purpose
Read or set the temperature calibration for the active channel
Syntax
“TCAL=
adder, mult, date
”
“TCAL”
Arguments
adder
:
mult
:
date
:
Temperature adder in the current temperature units.
Temperature multiplier.
Date (YYYYMMDD)
Remarks
The temperature calibration for active channel is applied to the displayed molbloc
temperature.
Example
Command:
Reply:
“TCAL=0.25, 1.003,20090226”
“0.250, 1.0030,20090226”
Errors
ERR# 2:
The
date
argument is invalid
See Also
3.2.2, “TUNIT”, “DEVICE”, “STDRES”
TEMP
Purpose
Read the temperature of the active molbloc in the current unit of measure.
Syntax
“TEMP”
Remarks
The molbloc temperature is returned in the current temperature units. The average
temperature, upstream temperature and downstream temperature is returned (in
that order). An error will be returned if there is a problem reading the molbloc
PRTs.
Example
Command:
Reply:
“TEMP”
“17.85 C, 17.84 C, 17.86 C”
Errors
ERR# 37:
There is an error in the PRT measurement.
See Also
3.2.5, 5.4, 3.4.5 “TUNIT”, “OHMS”, “TCAL”
TIME(=)
Purpose
Read or set the internal clock time.
Syntax
“TIME”
“TIME=hh:mm:m
Arguments
hh:mmxs:
The time to set in the format where hh is the hours from 1 to 12,
mm is the minutes from 1 to 59 followed by “am” or “pm”.
Example
Command:
Reply:
“TIME=1:22am”
“01:22am”
Errors
ERR# 7:
The time given is invalid.
See Also
3.6.5.3