Test Result Format
Ex: When LCR Meter have Processed Measurement, then will sequent send 2 command for test result automatically. Refer to below:
Ex: When C=1nF,D= .0045 (C/D mode and display mode is VALUE)
1. Primary factor of test result. (It doesn’t include primary's unit).
LCR Meter response MAIN:PRIM 1.0000^END
ASCII CODE = 4D 41 49 4E 3A 50 52 49 4D 20 20 31 2E 30 30 30 30 0A (Hex format)
Positive symbol
2. Secondary factor and primary’s unit
LCR Meter response MAIN:SECO .0045nF^END
ASCII CODE = 4D 41 49 4E 3A 53 45 43 4F 20 20 2E 30 30 34 35 6E 46 0A (Hex format)
Positive symbol Primary factor’ unit
Ex: When R=1ohm,Q= .0005 (R/Q mode and display mode is VALUE)
1. Primary factor of test result (It doesn’t include primary's unit.
LCR Meter response MAIN:PRIM 1.0000^END
ASCII CODE = 4D 41 49 4E 3A 50 52 49 4D 20 20 31 2E 30 30 30 30 0A (Hex format)
Positive symbol
2. Secondary factor and unit
LCR Meter response MAIN:SECO .0005 ^END
ASCII CODE = 4D 41 49 4E 3A 53 45 43 4F 20 20 2E 30 30 30 35 20 20 0A (Hex format)
Positive symbol Primary factor’s unit