Constant voltage and constant current
If the output load resistance exceeds the voltage setting divided by the current setting, the instrument will operate
in constant-voltage (CV) mode. The current will equal the voltage divided by the load resistance.
If the output load resistance is less than the voltage setting divided by the current setting, the instrument will
operate in constant-current (CC) mode. The voltage will equal the current multiplied by the load resistance.
From the remote interface:
A channel parameter is required with each SCPI command to select an output. For example, (@1) selects output 1,
(@2,3) selects output 2 and 3, and (@1:3) selects outputs 1 through 3. The output list must be preceded with an @
symbol and be enclosed in parentheses ().
To set only output 1 to 5 V and 1 A:
APPL Ch1 5, 1
To enable only output 1:
OUTP ON,(@1)
To enable output 1 and output 3:
OUTP ON,(@1,3)
To measure the output voltage and current of output 1:
MEAS:VOLT? (@1)
MEAS:CURR? (@1)
58
Keysight E36300 Series User's Guide