Robot Command Interface
0112-0151 K
149
Ren
Action
:
Renames <index> plate to <name>, given in double quotes
and limited to 19 characters.
Ren
Response
:
“No data OK” response if the command succeeds.
Del
Action
:
Deletes <index> plate.
Del
Response
:
“No data OK” response if the command succeeds.
Get
Action
:
Reads the <index> plate description parameters.
Get
Response
:
Returns the parameters as a list of single character tokens
followed by the integer token value, where the tokens are:
W:
Plate well type: 0 = 96 wells, 1 = 384 wells.
B:
Bottom shape: 0 = flat, 1 = U-bottom, 2 = V-bottom, 3
= small volume.
D:
Well depth, microns (0-plate height).
H:
Plate height, microns (7500-25500).
Get
Example
:
/plate get 2
CR
returns OK
TAB
W0 B0 D14400 H10900
CRLF
>
Set
Action
:
Sets the <index> plate description parameters to the values
given in <Args>. <Args> is a list of single character tokens
followed by the integer token value, where the tokens are:
W:
Plate well type: 0 = 96 wells, 1 = 384 wells.
B:
Bottom shape: 0 = flat, 1 = U-bottom, 2 = V-bottom, 3
= small volume.
D:
Well depth, microns.
H:
Plate height, microns.
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...