Example 4:
@123V00002
Acknowledge 4:
@123 Load A 120.45 Lb
@123 Load A 125.29 Lb
.....
@123 Load A 137.81 Lb
Print Command (P)
The print command will fetch a full set of readings (like pressing the front panel print button). These will be
in the units that are currently selected on the numerical display. The format is P(repeat number), where the
repeat number must be 0, 1, 2, or 3. Repeat 1 will return 1 set of readings to the RS232 terminal. Repeat 2 will
return a set of readings approximately every three seconds (like the value command). Repeat 0 will turn off
printing that was started with repeat 2. Repeat 3 will send one set of readings to the serial printer attached to
the Model 4215-HS instead of to the RS232 terminal.
Example:
@123P1
Acknowledge:
@123 Load A 45.625 Lb
Peak A 82.396 Lb Vall A -0.005 Lb
Load B 22.560 Lb
Peak B 23.000 Lb
Vall B –0.0001 Lb
Pos 3.500 In
Vel 0.0 I/M
Avg A 0.000 Lb
Avg B 0.000 Lb
Limits 0 0 0 1
Reset Tare, Peak, Valley, or Position Command (R)
This command allows you to reset the tare, peak, valley, or position readings. The format is R(tare A)(peak A)
(valley A) (tare B)(peak B)(valley B) (position select), where the selects are either a 0 (don’t reset) or a 1 (reset).
Thus you can reset any combination of these readings with a single command.
Example:
@123R1000000
Acknowledge:
@123 Reset – Tare A