NI 4065 Calibration Procedure
52
ni.com
c.
Call the niDMM Control VI.
d.
On the calibrator, output the value listed in the
Calibrator Output
column in Table 14 for the current iteration.
Allow the calibrator
output to settle before proceeding
.
e.
Call the niDMM Read VI.
f.
Verify that this measurement falls between the limits listed in
Table 26.
8.
Repeat step 7 for each of the remaining iterations shown in Table 14.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_Control
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Control Action
:
Commit
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Reading
: Verify that this measurement
falls between the limits listed in Table 26
Maximum_Time
: –
1
Table 14.
AC Current Settings
Iteration
Range (A
ac
)
Calibrator Output
Amplitude (A)
Frequency
1
0.01
200
1 kHz
2
0.01
1 m
1 kHz
3
0.01
10 m
1 kHz
4
0.1
2 m
1 kHz
5
0.1
10 m
1 kHz
6
0.1
100 m
1 kHz
7
0.5
10 m
1 kHz