National Instruments Corporation
69
NI 4065 Calibration Procedure
23. Call the niDMM Cal Adjust Gain VI.
24. Output 100
from the calibrator with 2-wire compensation and
external sense turned on.
Allow the calibrator output to settle before
proceeding
.
25. Call the niDMM Cal Adjust Gain VI.
26. Output 0
from the calibrator with 2-wire compensation turned off,
but with external sense turned on.
Allow the calibrator output to settle
before proceeding
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
1 k
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 1 k
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 100