NI 4065 Calibration Procedure
11. Call the niDMM Cal Adjust Offset VI.
12. Output 0.1 A on the calibrator.
Allow the calibrator output to settle
before proceeding
.
13. Call the niDMM Cal Adjust Gain VI.
14. Output –0.1 A on the calibrator.
Allow the calibrator output to settle
before proceeding
.
15. Call the niDMM Cal Adjust Gain VI.
16. Output 0 A on the calibrator.
Allow the calibrator output to settle
before proceeding
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
0.01
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
0.1
I
nput_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
0.1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
0.1
I
nput_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
–
0.1