NI 4065 Calibration Procedure
40
ni.com
f.
Call the niDMM Read VI.
•
For iterations containing the 0
measurement of each range,
store the result as the offset null for that range.
•
For all other iterations, subtract the previously stored offset
null for the corresponding range from the measurement taken.
g.
Verify that this measurement falls between the tolerances listed in
Table 23. Tolerances are provided instead of absolute limits
because your calibrator will have different discrete resistance
values.
5.
Repeat step 4 for each of the remaining iterations shown in Table 10.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Reading
: Verify that this measurement
falls between the tolerances listed in
Table 23
Maximum_Time
: –
1
Table 10.
2-Wire Resistance Settings
Iteration
Range (
)
Calibrator Output
Resistance
(
Delay
(seconds)
2-Wire
Compensation
External
Sense
1
100
0
0.0
ON
ON
2
10
0.0
ON
ON
3
100
0.0
ON
ON
4
1 k
0
0.0
ON
ON
5
100
0.0
ON
ON
6
1 k
0.0
ON
ON
7
10 k
0
0.0
ON
ON
8
1 k
0.0
ON
ON
9
10 k
0.1
ON
ON