Keyword
Parameter
Std/
New
Comments
[:PULSe] [0]
Std
Subsystem. Contains commands to control
the output pulse generation. Commands
without suffix refer to the currently selected
logical instrument. See INSTrument
subsystem.
:STATe
<boolean value>
Std
Enables / Disables the output for all channels.
Command is the same as pressing the
RUN/STOP button.
:PERiod
<numeric value>
Std
Sets the To period.
:MODe
NORMal /
SINGle /
BURSt /
DCYCle
New
Sets the To mode.
:BCOunter
<numeric value>
New
Burst Counter. Number of pulses to generate
in the Burst mode.
:PCOunter
<numeric value>
New
Pulse Counter. Number of pulses to generate
during on cycle of the Duty Cycle mode.
:OCOunter
<numeric value>
New
Off Counter. Number of pulses to inhibit output
during the off cycle of the Duty Cycle mode.
:CLock
SYS/
EXT10/
EXT20/
EXT25/
EXT40/
EXT50/
EXT80/
EXT100
New
Sets Source for the external rate generator.
System Clock or External Source ranging from
10MHz to 100MHz.
:OCLock
To / 10 / 11 / 12 /
14 / 16 / 20 / 25 /
33 / 50 / 100
New
Sets external clock output. To Pulse or 50%
duty cycle TTL output from 10MHz to 100MHz.
:GATe
New
Subsystem. Contains the commands to define
the Gate function.
:MODe
DIS/PULS/OUTP/
CHAN
New
Sets Global Gate Mode. Disable, pulse
inhibit, output inhibit, channel.
:LOGic
LOW / HIGH
New
Sets Channel Gate logic level. Active low or
active high.
:LEVel
<numeric value>
New
Sets the gate threshold. Value is in volts with
a range of .20 to 15 Volts.
:TRIGger
New
Subsystem. Contains the commands to define
the Trigger function.
:MODe
DIS / TRIG
New
Sets Trigger Mode. Disable or TRIG (enable).
:EDGe
RISing / FALLing
New
Selects which edge (rising or falling) to use as
the trigger signal.
:LEVel
<numeric value>
New
Sets the Trigger Threshold. Value is in volts,
with a a range of .20 to 15 Volts.
9520 SCPI Command Summary