Name
Type
Range
Default
<bool>
Bool
{0|1|ON|OFF}
0
Remarks
•
Before enabling the channel output, please make sure that the current setting
will not affect the devices connected to the power supply.
•
If <source> is omitted, the command executes the corresponding operation on
the current channel.
Return Format
The query returns 1 or 0.
Examples
:OUTP CH1,ON /*Enables the output for CH1.*/
:OUTP? CH1 /*Queries whether the CH1 output is enabled. The query
returns 1*/
4.8.16
:OUTPut:TRACk[:STATe]
Syntax
:OUTPut:TRACk[:STATe]
<
bool>
:OUTPut:TRACk[:STATe]?
Description
Sets or queries the on/off state of tracking function.
Parameter
Name
Type
Range
Default
<bool>
Bool
{0|1|ON|OFF}
0
Remarks
•
For the two channels (CH1 and CH2) that support this mode, changes made on
one channel (voltage/current setting value, OVP/OCP level and on/off status) are
applied to the other channel.
•
By default, the tracking function is disabled. It is usually used to provide
symmetric voltage for the operational amplifier or other circuits.
•
The tracking function only tracks the voltage/current setting value. The actual
output voltage/current will not be affected.
Return Format
The query returns 1 or 0.
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
57