Return Format
The query returns 1 or 0.
Examples
:CURR:PROT:TRIP? /*Queries whether an OCP event occurred on the
current channel. The query returns 1.*/
4.9.7
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate]
[:AMPLitude]
Syntax
[:SOURce
[<
n>]
]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
{<
voltage>|
<
lim>|<amp>}
[:SOURce
[<
n>]
]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
[<
lim>]
Description
Sets or queries the voltage for the specified channel.
Parameter
Name
Type
Range
Default
<n>
Integer
{1|2|3}
-
<voltage>
Real
Please refer to
Table 4.33 Range and default value
of overvoltage/overcurrent protection
<lim>
Discrete
{MINimum|MAXimum|DEFault}
-
<amp>
Discrete
{UP|DOWN}
-
Remarks
•
If [:SOURce[<n>]] or [<n>] is omitted, the command sets the corresponding
parameter for the channel currently selected.
•
When <voltage> is selected, the command directly sets the voltage for the
specified channel. When MINimum, MAXimum, or DEFault is selected, the
command sets the minimum, maximum, or default voltage value within the
available range for the specified channel. When UP or DOWN is selected, the
command steps up or down the voltage by the step size set in
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
.
•
to set the voltage and current for the specified
channel.
Command System
64
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Summary of Contents for DP900 Series
Page 1: ......
Page 133: ...d Exit Programming Examples Copyright RIGOL TECHNOLOGIES CO LTD All rights reserved 125 ...
Page 141: ......