Installation & Operation Manual for the
15
Active Vibration Cancellation (AVC) System
23)
Display the temperature sensor reading in kelvin
a)
Command:
TC<CR>
b)
Returns the cold head temperature in kelvin.
24)
Display the reject temperature in Celsius
a)
Command:
TEMP<SP>RJ<CR>
b)
Returns the current reject temperature measured at the base of the fins on the cooler. Reject temperature is displayed
in Celsius.
25)
Display target temperature in kelvin
a)
Command:
TTARGET<CR>
b)
Returns the target temperature of the controller in kelvin when in temperature control mode (default mode).
26)
Set target temperature
a)
Command:
TTARGET=<VAL><CR>
b)
This command is User locked.
c)
Set the temperature the cryocooler will try to attain (to ±0.1 K) when in temperature control mode.
d)
<VAL>
is the value of the desired target temperature in kelvin.
27)
Display current controller code version
a)
Command:
VERSION<CR>
b)
Returns the current controller coder version.
TEMP RJ
025.30
TC
295.21
TTARGET
077.00
TTARGET=77
077.00
VERSION
1.0.4