VTI Instruments Corporation
38
VM2710A Programming
MFGTEST? 6
Reports the offset value. The equvalent command is
MFGTEST:CAL_OFFSET?.
MFGTEST 7 #S
Sets the converter scale value for non-vol, where #S
is the value to be set.
MFGTEST? 7
Reports the converter scale value.
MFGTEST 8 #O
Sets the converter offset value for non-vol, where #O
is the value to be set.
MFGTEST? 8
Reports the converter offset value.
MFGTEST 10 #?
Sets the readings per trigger.
MFGTEST 10 #L #C
Sets a linearization reference point.
MFGTEST? 10 #L
Reports the linearization reference pint value.
MFGTEST 11 #L #C
Sets a linearization slope. There are 16 linearization
constants (0 - 15) for each aperture setting. #L
indicates which linearization segment is being set,
and #C is the constant to be loaded for the selected
linearization segment in the set aperture.
MFGTEST? 11 #L
Reports the linearization slope value.
MFGTEST 12 #L #C
Sets a linearization offset. There are 16 linearization
constants (0 - 15) for each aperture setting. #L
indicates which linearization segment is being set,
and #C is the constant to be loaded for the selected
linearization segment in the set aperture.
MFGTEST? 12 #L
Reports the linearization offset value.
MFGTEST 15 #count
Increment or decrement the gain slope by #count,
where #count is a positive (increment) or negative
(decrement) integer. See MFGTEST 5.
MFGTEST 16 #count
Increment or decrement the offset by #count, where
#count is a positive (increment) or negative
(decrement) integer. See MFGTEST 6.
MFGTEST 17 #count
Increment or decrement the converter scale value by
#count, where #count is a positive (increment) or
negative (decrement) integer. See MFGTEST 7.
MFGTEST 18 #count
Increment or decrement the converter offset value by
#count, where #count is a positive (increment) or
negative (decrement) integer. See MFGTEST 8.