www.vtiinstruments.com
VM2710A Command Dictionary
153
TRIGger:COUNt
Purpose
Sets the number of triggers
Type
Setting
Command Syntax
TRIGger:COUNt <number>
Command Parameters
<number> = 0 through 16,777,215 | MIN | MAX|CONTinuous
*RST Value
1
Query Syntax
TRIGger:COUNt? [<MIN | MAX>]
Query Parameters
<MIN> = Causes the query to report the minimum value (1)
<MAX> = Causes the query to report the maximum value (16,777,215)
Query Response
0 through 16,777,215
Description
This command sets the number of triggers issued. When setting the trigger count, the
memory (64 k or 256 k readings) must be taken into consideration, for the trigger count
multiplied by the sample count (SAMPle:COUNt) must be less than or equal to the
memory capability.
When set to 0 (countinuous) the voltmeter returns to the waiting for trigger state after
each trigger. This is particularly useful when using the register interface and an extrenal
trigger. Once INITiate is used to start operation, ABORt or *RST command must be
used to remove the voltmeter form the countinuous opeation.
The CONFigure and MEASure commands set the trigger count to 1.
Examples
Command / Query
Response (Description)
TRIG:COUN 10
TRIG:COUN?
10
Related Commands
SAMPle:COUNt