VTI Instruments Corporation
148
VM2710A Command Dictionary
[SENSe:]VOLTage:RESolution:AC
Purpose
Selects the resolution for ac voltage measurements
Type
Setting
Command Syntax
[SENSe:]VOLTage:RESolution:AC <resolution>
Command Parameters
<resolution> = 0.0000001 (100 nV)
0.000001 (1
µ
V)
0.00001 (10
µ
V)
0.0001 (100
µ
V)
0.001 (1 mV)
0.01 (10 mV)
DEF | MIN | MAX
*RST Value
Autoranging
Query Syntax
[SENSe:]VOLTage:RESolution:AC? <MIN | MAX>
Query Parameters
<MIN> = Causes the query to report the minimum value (0.0000001)
<MAX> = Causes the query to report the maximum value (0.01)
Query Response
0.0000001 | 0.000001 | 0.00001 | 0.0001 | 0.001 | 0.01
Description
This command sets the resolution for voltage measurements. MINimum sets the value
to 100 nV; MAXimum sets the value to 10 mV. Note that exponential form may also
be used for the resolution value.
Setting resolution also sets the aperture time and the integration time in power line
cycles. This command can override the previous values set by VOLTage:APERture or
VOLTage:NPLC. The last command executed has priority.
Examples
Command / Query
Response (Description)
VOLT:RES 0.001
(Sets the resolution to 1 mV.)
Related Commands
[SENSe:]VOLTage:APERture
[SENSe:]VOLtage:NPLC