60
Chapter 1
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Questionable Integrity Positive Transition
This command determines which bits in the
Questionable Integrity Condition register will set the corresponding bit in the
Questionable Integrity Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Questionable Integrity Signal Register
Questionable Integrity Signal Condition
This query returns the decimal value of the sum of
the bits in the Questionable Integrity Signal Condition register.
NOTE
The data in this register is continuously updated and reflects the
current conditions.
Example
STAT:QUES:INT:NTR 8 Measurement ‘regained calibration’
Summary will be reported to the Integrity Summary of the
Status Questionable register.
Preset
0
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:PTRansition
<integer>
:STATus:QUEStionable:INTegrity:PTRansition?
Example
STAT:QUES:INT:PTR 8 Measurement ‘became uncalibrated’
Summary will be reported to the Integrity Summary of the
Status Questionable register.
Preset
32767
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:SIGNal:CONDition?
Example
STAT:QUES:INT:SIGN:COND?
Preset
0