Page 37
VRANGE<
N
>?
Returns the voltage range for output<
N
>. The response is
<
NR
1><
RMT
> where <
NR
1> has the following meaning:
Output1: 0= Disable, 1= 35V/3A, 2 = 16V/6A, 3 = 35V/6A.
Output2: 0= Disable, 1= 35V/3A, 2 = 16V/6A, 3 = 35V/6A.
Output3: 0= Disable, 1= 35V/3A, 2 = 70V/1.5A, 3 = 70V/3A.
Output4: 0= Disable, 1= 35V/3A, 2 = 70V/1.5A, 3 = 70V/3A.
CONFIG <
NRF
>
Sets the voltage tracking mode of the unit to <
NRF
> where <
NRF
>
has the following meaning:
0 = None.
1 = Mode1.
2 = Mode2.
3 = Mode3.
These modes are as defined within the Setting Voltage Tracking
section of this manual – see section 8.1.
CONFIG?
Returns the voltage tracking mode of the unit. The response is
<
NR
1><
RMT
>, where <
NR
1> has the following meaning:
0 = None.
1 = Mode1.
2 = Mode2.
3 = Mode3.
These modes are as defined within the Setting Voltage Tracking
section of this manual – see section 8.1.
ONDELAY<
N
> <
NRF
>
Set output<
N
> Multi-On delay where <
NRF
> is in milliseconds
OFFDELAY<
N
> <
NRF
> Set output<
N
> Multi-Off delay where <
NRF
> is in milliseconds
ONACTION<
N
> <
CPD
> Set output<
N
> Multi-On action where <
CPD
> can be QUICK, NEVER
or DELAY.
OFFACTION<
N
> <
CPD
> Set output<
N
> Multi-Off action where <
CPD
> can be QUICK, NEVER
or DELAY.
SAV<
N
> <
NRF
>
Save the current settings of output<
N
> to the store specified by
<
NRF
> where <
NRF
> can be 0-49.
RCL<
N
> <
NRF
>
Recall the settings for output<
N
> from the store specified by <
NRF
>
where <
NRF
> can be 0-49.
12.2.2 Common Commands
*IDN?
Returns the instrument identification.
The response is in the form <NAME>, <model>, <serial>, <version><
RMT
>
where <NAME> is the manufacturer's name, <model> is the instrument
type, <serial> is the interface serial number and <version> is the revision
level of the firmware installed.
*RST
Resets the functional parameters of the instrument to the default settings as
listed in the Factory Default Settings, see section
Error! Reference source
not found.
.
Does not affect the contents of the Save and Recall stores.
Does not affect any remote interface settings.
*SAV <
NRF
>
Save the settings for all four outputs simultaneously to the store specified by
<
NRF
> where <
NRF
> can be 0-49. This includes output On/Off state, current
meter averaging state, and the Multi-On/Multi-Off settings.