Standard Commands
(1) System Data Command
Query Instrument ID (Identification Code)
Syntax
Query
*IDN?
Response
<Manufacturer name>,<Model name>,<Serial number>,<Software version>
Example
*IDN?
HIOKI, SM7110,123456,V1.00
The Instrument ID is HIOKI SM7110, serial number 123456, software version 1.00.
(2) Internal Operation Command
Execute Self-Test and Query Result
Syntax
Query
*TST?
Response
<0 to 1 (NR1)>
<0> = No error
<1> = ROM error
Description
Perform the instrument self-test and return the result as NR1 value 0 or 1.
Returns zero when no error occurs.
Example
*TST?
1
A ROM error occurred.
(3) Synchronization Commands
Set OPC bit of SESR when Finished with All Pending Operations
Syntax
Command
*OPC
Description
Sets
OPC
bit 0 of the Standard Event Status Register (SESR) when all commands prior
to
*OPC
have finished processing.
Example
A;B;*OPC;C
The OPC bit of SESR is set after A and B command processing has been completed.
Respond with ASCII “1” when Finished with All Pending Operations
Syntax
Query
*OPC?
Response
1
Description
Response with ASCII “1” when all commands prior to
*OPC
have finished processing.
Initialize Instrument (Normal Reset)
Syntax
Command
*RST
Description
Command
Resets the instrument to factory settings excluding the communication state, ,
zero adjustment value and saved data.(Normal Reset)
The initial display is displayed after initialization.
Note
The communication state is not initialized.
Use the
:RESet:SYSTem
command to initialize the zero adjustment value
and saved data.