22
(10)
Range
Set and Query Current Range
Syntax
command
:RANGe
<20pA/200pA/2nA/20nA/200nA/2uA/20uA/200uA/2mA>
Query
:RANGe?
Response
<20pA/200pA/2nA/20nA/200nA/2uA/20uA/200uA/2mA>
Example
:RANGe
20pA
Set current range to 20pA.
:RANGe?
20pA
Current range has been set to 20pA.
Set and Query Auto Range
Syntax
command
:RANGe:AUTO
<ON/OFF>
Query
:RANGe:AUTO?
Response
<ON/OFF>
Description
ON … Auto range is ON.
OFF … Auto range is OFF ( HOLD range )
Example
:RANGe:AUTO
ON
Set auto range to ON.
:RANGe:AUTO?
ON
Auto range has been set to ON.
(11)
Delay
Set and Query Delay time
Syntax
command
:DELay
<Delay time(s)(NR2)>
Query
:DELay?
Response
<Delay time(s)(NR2)>
Description
<Delay time(s)(NR2)> = 0.0
~
999.9
Example
:DELay 12.3
Set delay time to 12.3 seconds
:DELay?
12.3
Delay time has been 12.3 seconds
(12)
Average
Set and Query Average
Syntax
command
:AVERage
<OFF/HOLD/AUTO>
Query
:AVERage?
Response
<OFF/HOLD/AUTO>
Description
OFF … Average is OFF
HOLD … Moving average
AUTO … The number of times of averaging is automatically changed based on
the variation of the measured values
Example
:AVERage HOLD
Set average to moving average
:AVERage?
HOLD
Average has been moving average
Summary of Contents for Super megohm SM7110