2/3
SRT
Function to interact with measuring voltage ON or measurement start. SRT
:STARt
STP
Function to interact with measuring stop (measuring voltage OFF).
STP
:STOP
MTG
Generates manual trigger.
MTG
*TRG
IVS
Setting Measuring Voltage .
IVS <0.1~1000.0>
:VOLTage <Voltage(V)(NR2)>
IVS?
Query response of Measuring Voltage is same as setting.
IVS?
:VOLTage?
PWS
Measuring power source setting.
PWS <0~2>,<0~1>,<0~1>
• Because the instrument does Not support the Noise filter will be igNored.
:CHARge:LIMit:CURRent <1.8mA/5mA/10mA/50mA>
:CHARge:LIMit <ON/OFF>
PWS?
Measuring power source setting query.
Query response in display mode is same as setting.
PWS?
• Because the instrument does Not support the Noise filter will always be "0."
:CHARge:LIMit:CURRent?
:CHARge:LIMit?
DEV
Deviation value display mode setting.
DEV <0~1>,<-9.9999E+30~
9.9999E+30>
• Because the instrument does Not support deviation value display mode, this command will be igNored and will Not
result in a command error.
Not compatible
DEV?
Query response in deviation value display mode is same as setting.
DEV?
• Because the instrument does Not support deviation value display mode, this command will be igNored and will Not
result in a command error.
Not compatible
ELC
Electrode data setting.
ELC <0~1>,<0.0~999.9>,<0.1
~1199.9>,<0.001~
30.000>,<0.01~999.99>
• Because the instrument does Not support SIZE/ACTL selection, the SIZE/ACTL setting will be igNored.
:ELECtric:D1 <diameter(m)(NR2)>
:ELECtric:D2 <diameter of theconter(m)(NR2)>
:ELECtric:T <Thickness (m)(NR2)>
:ELECtric:K <Electrode constact(NR2)>
ELC?
Query response electrode data is same as setting.
ELC?
• Because the instrument does Not support SIZE/ACTL selection, the SIZE/ACTL response will always be "0."
:ELECtric:D1?
:ELECtric:D2?
:ELECtric:T?
:ELECtric:K?
VCM
Selection of auto voltage check execute mode.
VCM <0~1>
:VCHeck:STATe <ON/OFF>
VCM?
Query response of auto voltage monitoring execute mode is same as
setting.
VCM?
:VCHeck:STATe?
CCM
Selection of auto contact check execute mode.
CCM <0~1>
:CONTactcheck:STATe <ON/OFF>
CCM?
Query response of auto contact check execute mode is same as
CCM?
:CONTactcheck:STATe?
WCP
Work capacity setting Capacity of work for open correction value
calculation.
WCP <0.5~99.9>
:CONTactcheck:LIMit <Judgment reference value (F)(NR3)>
WCP?
Query response of work capacity is same as setting.
WCP?
:CONTactcheck:LIMit?
OST?
Executes open correction once and capacity value measured during
open correction is returned as a response.
OST?
• The response range is 0 to 99.999.
• In the event of an error, the query will result in a response of 999.9.
:OPEN?
:OPEN:VALue?
VCK?
Executes voltage check once and returns result of it as a response.
VCK?
:VCHeck?
CCK?
Executes contact check once and returns result of it as a response.
CCK?
:CONTactcheck?
TGM
Trigger mode setting.
TGM <0~2>
• Because the instrument does Not support the "1: Manual trigger" setting, specifying "1" will cause an external
trigger to be set.
:TRIGger <INTernal/EXTernal>
TGM?
Query response of trigger mode is same as setting.
TGM?
• Because the instrument does Not support the "1" setting, the response will be either "0" or "2."
:TRIGger?
RDT?
Measured data query.
RDT? <0~2>
:MEASure?
:MEASure:COMParator?
:MEASure:RESult? <1~255>
CMP
Comparison Measuring mode setting.
CMP <0~1>,<0~2>,<-
9.9999E+30~9.9999E+30>,<-
9.9999E+30~9.9999E+30>
• Because the instrument does Not support mode selection, the "Mode" setting will be igNored.
:COMParator:LIMit <OFF/Upper-limit(NR3)>,<OFF/Lower-limit(NR3)>
CMP?
Query response of Comparison Measuring mode is same as setting.
CMP?
• The "Comparison execution" query will result in a response of "1" if either the upper limit value or lower limit
value is enabled.
• The "Mode" query will always result in a response of "1."
:COMParator:LIMit?
SEQ
Sequential mode setting.
SEQ <0~1>,<0~9>,<0.0~
999.9>,<0.0~999.9>,<0.0~
999.9>,<0.0~999.9>
:SEQuence:STATe <ON/OFF>
:SEQuence:NUMBer <0~9(NR1)>
:SEQuence:TIME <Sequential Measuring No.(NR1)>,<Discharge time
before measurement starts(s)(NR2)>,<Charging (voltage applied) time(s)
(NR2)>,<Measuring Time(s) (NR2)>,<Discharge time after completion of
measurement(s) (NR2)>
SEQ?
Same as set items of query SEQ command message in sequential
mode.
SEQ?
:SEQuence:STATe?
:SEQuence:NUMBer?
:SEQuence:TIME? <Sequential Measuring No.(NR1)>
CNF
Operation environment setting.
CNF <0~1>,<0~1>,<0~1>,<0
~1>
• Specifying "0" for the "Beep tone" parameter will disable all comparator judgment tones.
• Specifying "1: for the "Beep tone" parameter and "0" for the "Beep tone PASS/FAIL" parameter will set the
comparator judgment tone for HI and LO judgments to Type 1 and while disabling the judgment tone for IN
judgments.
• Specifying "1" for both the "Beep tone" parameter and the "Beep tone PASS/FAIL" parameter will disable the
comparator judgment tone for HI and LO judgments while setting the tone for IN judgments to Type 1.
:INTerlock <ON/OFF>
:COMParator:BEEPer
<HI/IN/LO>,<TYPE1/TYPE2/TYPE3/OFF>,<Times>
:KEY:BEEPer <ON/OFF>
CNF?
Query response of operation environment is same as setting.
CNF?
• "Beep tone" queries will result in a response of "0" when all comparator judgment tones are disabled.
• "Beep tone PASS/FAIL" queries will result in a response of "1" when the comparator judgment tone for IN
judgments is Not disabled.
:INTerlock?
:COMParator:BEEPer? <HI/IN/LO>
:KEY:BEEPer?
THL
Histogram display threshold setting.
• The instrument does Not provide a histogram display.This command will be igNored and will Not result in a
command error.
Not compatible
THL?
Query response of threshold is same as setting.
• The instrument does Not provide a histogram display.This command will be igNored and will Not result in a
command error.
Not compatible
RHS?
Histogram counter readout.
• The instrument does Not provide a histogram display.This command will be igNored and will Not result in a
command error.
Not compatible
CHS
Histogram counter clear.
• The instrument does Not provide a histogram display.This command will be igNored and will Not result in a
command error.
Not compatible