39
Set and Query TRIG Signal's Filter Time
Syntax
command
:IO:FILTer:TIME
< Filter time (s)(NR2)>
Query
:IO:FILTer:TIME?
Response
< Filter time (s)(NR2)>
Description
<
Filter time (s)(NR2)> = 0.001
~
0.500
Example
:IO:FILTer:TIME 0.123
Set trigger filter time to 0.133 seconds.
:IO:FILTer:TIME?
0.123
Trigger filter time has been set to 0.123 seconds.
Set and Query
GO-signal outputting logic level
Syntax
command
:IO:GOLogic
<NORMal/ INVert >
Query
:IO:GOLogic?
Response
<NORMAL/INVERT>
Description
NORMAL …When a pass judgment is given for the contact check or
voltage monitor check: Outputs a low-level signal.
When a fail judgment is given for the contact check or
voltage monitor check: Outputs a high-level signal.
INVERT …When a pass judgment is given for the contact check or
voltage monitor check: Outputs a high-level signal.
When a fail judgment is given for the contact check or
voltage monitor check: Outputs a low-level signal.
Example
:IO:GOLogic NORMal
Set GO-signal outputting logic level to NORMAL.
:IO:GOLogic?
NORMAL
GO-signal outputting logic level has been set to NORMAL.
Set and Query EOM Output Method
Syntax
command
:IO:EOM:MODE
<HOLD/PULSe>
Query
:IO:EOM:MODE?
Response
<HOLD/PULSe>
Description
HOLD … Holds the EOM signal until measurement starts by the next trigger signal.
PULSE … Sets EOM=OFF according to the specified pulse width
Example
:IO:EOM:MODE HOLD
Set EOM output method to HOLD.
:IO:EOM:MODE?
HOLD
EOM output method has been set to HOLD.
Set and Query EOM Pulse Width
Syntax
command
:IO:EOM:PULSe
< Pulse width (s)(NR2)>
Query
:IO:EOM:PULSe?
Response
< Pulse width (s)(NR2)>
Description
<
Pulse width (s)(NR2)> = 0.001
~
0.100
Example
:IO:EOM:PULSe 0.023
Set Pulse width to 0.023 seconds.
:IO:EOM:PULSe?
0.23
Pulse width has been set to 0.023 seconds.
Summary of Contents for Super megohm SM7110