30
Set and Query contact check state
Syntax
command
:CONTactcheck:STATe
<ON/OFF>
Query
:CONTactcheck:STATe?
Response
<ON/OFF>
Description
ON … Auto execute before measurement
OFF … No execute
Example
:CONTactcheck:STATe
ON
Set contact check auto execute to ON
:CONTactcheck:STATe?
ON
Contact check auto execute has been ON
Set and Query contact check limit
Syntax
command
:CONTactcheck:LIMit
< Judgment reference value(NR3)>
Query
:CONTactcheck:LIMit?
Response
< Judgment reference value (F)(NR3)>
Description
< Judgment reference value (NR3)> = 0.00
~
99.99E-12
Example
:CONTactcheck:LIMit 50E-12
Set contact check limit to 50pF
:CONTactcheck:LIMit?
50.00E-12
Contact check limit has been 50pF
Execute and Query contact check
Syntax
Query
:CONTactcheck?
Response
<0
~
1(NR1)>
Description
Execute contact check, retuen judgment
0 … contact check NG
1 …contact check OK
Example
:CONTactcheck?
1
Contact check has been OK
Query contact check
value
Syntax
Query
:CONTactcheck:VALue?
Response
<contact check value(NR3)>
Description
Retuen contact check value
Unexecuted jig capacity open correction … 99.999E-12
Unexecuted contact check … 99.999E-12
Example
:CONTactcheck:VALue?
+1.301311E-14
Contact check value has been 1.301311E-14F
Summary of Contents for Super megohm SM7110