SCPI Reference Tables
14-7
Table 14-7
SENSe command summary
Command
Description
Default
Parameter
Ref
SCPI
:SENSe[1]
:FUNCtion <name>
Select function; ‘VOLTage[:DC]’ or ‘TEMPerature’. ‘VOLT’
√
:FUNCtion?
Query measurement function.
√
:DATA
Path to return instrument readings:
√
[:LATest]?
Return the last instrument reading.
√
:FRESh?
Return a new (fresh) reading.
√
:CHANnel <n>
Select channel to measure; 0, 1 or 2 (0 = internal
temperature sensor).
1
:CHANnel?
Query selected channel.
:HOLD
Path to configure and control Hold:
:WINDow <n>
Set Hold window; 0.01 to 20 (%).
1
:WINDow?
Query Hold window.
:COUNt <n>
Set Hold count; 2 to 100.
5
:COUNt?
Query Hold count.
:STATe <b>
Enable or disable Hold.
OFF
:STATe?
Query state of Hold.
:VOLTage[:DC]
Path to configure DC volts:
√
:NPLCycles <n>
Set integration rate in line cycles (PLC); 0.01 to 60
(60Hz) or 0.01 to 50 (50Hz).
5
√
:NPLCycles?
Query NPLC.
√
:APERture <n>
Set integration rate in seconds; 166.67e–6 to 1
(60Hz) or 200e–6 to 1 (50Hz).
83.33
:APERture?
Query Aperture.
:DIGits <n>
Set display resolution; 4 to 8.
8
:DIGits?
Query display resolution.
:RATio <b>
Enable or disable Ratio; V1/V2.
OFF
:RATio?
Query state of Ratio.
:DELTa <b>
Enable or disable Delta; (V1t1 – V1t2)/2.
OFF
:DELTa?
Query state of Delta.
[:CHANnel1]
Channel 1 voltage commands:
:RANGe
Configure measurement range:
√
[:UPPer] <n>
Select range; 0 to 120 (volts).
120
√
[:UPPer]?
Query range value.
√
:AUTO <b>
Enable or disable autorange.
ON
√
:AUTO?
Query state of autorange.
√
:REFerence <n>
Specify reference (Rel) value for Channel 1;
–120 to 120 (volts).
0
√
:STATe <b>
Enable or disable Rel.
OFF
√
:STATe?
Query state of Rel.
√
:ACQuire
Use input signal as Rel.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...