Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 68 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.11
Summary of SCPI Commands
The following conventions are used in the SCPI command syntax:
•
Optional key words or parameters are issued in square brackets [ ]
•
Parameters within a command character string are issued in braces { }
•
The parameter issued in angle brackets < > must be replaced by a value.
9.11.1
Control Commands
*CLS
Reset all event registers
*ESE?
Query the event release register
*ESR[?]
Set or query the standard event register
*IDN?
Query identification
*OPC?
Return "1" for synchronization
*OPC
Set the event bit "Operation terminated"
*RST
Reset the gaussmeter
*SRE[?]
Set or query the sum release register
*STB?
Query the sum register
:STAT:PRES
Reset the error byte release register
:STAT:QUES:ENAB[?]
Set or query the error data event release
register
:STAT:QUES:EVEN?
Query the error data event register
:STAT:MEAS:ENAB[?]
Set or query the measurement event release
register
:STAT:MEAS:EVEN?
Query the measurement event register
9.11.2
Main Commands
:MEAS?
Query the current measuring value
:READ?
Query the current measuring value
:UNIT
{
TESL|APM|GAUS|OE|G|
T|O|A
}
:UNIT?
Set or query the physical unit
:MODE {DC|AC|PEAK}
:MODE?
Set or query the DC or AC field operating
mode
:RANG:SET {0|1|2|3}
Set the measuring range
:RANG?
Query the current measuring range
:NULL
Null compensation of the measuring probe