Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 87 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.12.7
Interface Functions
9.12.7.1
:SERI:COMA:BAUD[?]
Description:
Set or query the bit rate of the EIA-232 interface.
Mode:
Command or query
Parameter:
{1200|2400|4800|9600|14K4|19K2|38K4}
Response:
like parameter
*RST
Value:
not relevant
Example:
send
:SERI:COMA:BAUD 19K2<LF>
send
:SERI:COMA:BAUD?<LF>
receive 19K2<CR><LF>
9.12.7.2
:SERI:COMA:STOP[?]
Description:
Set or query the number of stop bits for the EIA-232 interface.
Mode:
Command or query
Parameter:
{1|2}
Response:
like parameter
*RST
Value:
not relevant
Example:
send
:SERI:COMA:STOP 2<LF>
send
:SERI:COMA:STOP?<LF>
receive 2<CR><LF>
9.12.7.3
:SERI:COMA:PARY[?]
Description:
Set or query the character parity of the EIA-232 interface.
Mode:
Command or query
Parameter:
{NONE|EVEN|ODD}
Response:
like parameter
*RST
Value:
not relevant
Example:
send
:SERI:COMA:PARY EVEN<LF>
send
:SERI:COMA:PARY?<LF>
receive EVEN<CR><LF>
Trial Version