24
Parameter
Rel.
Index
FF
Description
Type Store Size
byte
RO /
R/W Min. Max. Default
CAL_POINT_HI_2
56
The high calibration value applied to sensor 2
The value from either a calibrator or a reference equip-
ment.
Float
SRC
4
R/W
10
38
CAL_ACTUAL_HI_2
57
Entering any value will force the device to automatically
measure and store the actual high point value. Must be
entered with the applied CAL_POINT_HI_2 value
Float
SRC
4
R/W
10
38
SENSOR_CAL_METHOD_2
58
Enables or disables the last sensor calibration for
sensor 2
103 = Factory trim standard (calibration values disabled)
104 = User trim standard (calibration values enabled)
Un-
signed
8
SRC
1
R/W
103
SENSOR_CAL_LOC_2
59
The last location of the calibrated sensor
OCTET_
STRING
SRC
32
R/W
» »
SENSOR_CAL_DATE_2
60
The last date on which the calibration was performed
7 * Un-
signed
8
SRC
7
R/W
0,0,0,0,
1,1,103
SENSOR_CAL_WHO_2
61
The name of the person responsible for the last sensor
calibration
OCTET_
STRING
SRC
32
R/W
» »