Name
Address
Type
R
W
Description
Span calibration
0103
H
execute
N
F
Calibrate the scale span. Ensure that the
calibration weight is set first.
Edit linearisation
0104
H
execute
N
F
Edit a linearisation point, takes the
linearisation point as an argument. Ensure
that the calibration weight is set first.
Clear linearisation
0105
H
execute
N
F
Clear a linearisation point, takes the
linearisation point to be cleared as an
argument
Direct zero calibration
0106
H
execute
N
F
Direct mV/V zero calibration, takes the
mV/V value * 10000 at zero as an
argument (1.0000 mV/V should be entered
as 10000).
Direct span calibration
0107
H
execute
N
F
Direct mV/V span calibration, takes the
mV/V value * 10000 at full scale relative
to zero as an argument (1.0000 mV/V
should be entered as 10000).
Zero mV/V value
0111
H
int32
A
N
mV/V value of the zero point * 10000.
Span weight
0112
H
int32
A
N
Weight used for the span calibration.
Span mV/V value
0113
H
int32
A
N
mV/V value of the span point relative to
the zero point * 10000.
Linearisation correction
weight
0114
H
execute
A
A
Execute with the linearisation point
number (0 - 9) to get the correction weight
at that linearisation point.
Linearisation percentage
0115
H
execute
A
A
Execute with the linearisation point
number (0 - 9) to get the percentage value
between zero and span for that
linearisation point.
Number of decimal places
0128
H
uint8
A
F
Set the number of decimal places of the
scale.
Scale units
0129
H
uint8
A
F
Values:
0 is none
1 is grams
2 is kilograms
3 is pounds
4 is tonnes
5 is ounces
6 is user.
Current Time/Date
0150
H
string
A
N
Read this register to get current
instrument date/time (e.g. 10/12/2016
18:30:10).
Date Format
0151
H
uint8
A
A
Write 0 for DD.MM.YY
1 for DD.MM.YYYY
2 for MM.DD.YY
3 for MM.DD.YYYY
4 for YY.MM.DD
5 for YYYY.MM.DD
RTC Day
0152
H
uint8
A
A
Read/Write current day (1..31)
RTC Month
0153
H
uint8
A
A
Read/Write current month (1..12)
RTC Year
0154
H
uint16
A
A
Read/Write current year (2000..2099))
RTC Hour
0155
H
uint8
A
A
Read/Write current hour (0..23)
RTC Minute
0156
H
uint8
A
A
Read/Write current minute (0..59)
82
C500-600-2.0.3
Summary of Contents for C500 series
Page 1: ...C500 Series Digital Indicator Reference Manual ...
Page 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Page 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Page 123: ...118 C500 600 2 0 3 ...
Page 182: ...177 C500 600 2 0 3 ...