(9) The measured value of the multimeter is Vm3, and the value displayed
by SPS is Vsps3.
Record the values of Vm3 and Vsps3.
(10) Write the following two commands to calibrate the voltage setting value
Format: CALibrate:SET:VOLTage point1,point2,Vm1,Vset1,Vm2,Vset2
(Eg: CALibrate:SET:VOLTage 0,35,0.3227,0.5,34.636,35)
Format: CALibrate:SET:VOLTage point2,point3,Vm2,Vset2,Vm3,Vset3
(Eg: CALibrate:SET:VOLTage 35,80,34.636,35,74.428,75)
Note: if Vm2 >= point2, point2 = Vm2 + 0.1
(11) Send command CALibrate:SAVE CH1,VSET to save the calibration
data.
(12) Write the following two commands to calibrate the voltage measured
value
Format:
CALibrate:MEAS:VOLTage point1,point2,Vsps1,Vm1,Vsps2,Vm2
(Eg: CALibrate:MEAS:VOLTage 0,35,1.531,0.3227,36.318,34.636)
Format:
CALibrate:MEAS:VOLTage point2,point3,Vsps2,Vm2,Vsps3,Vm3
(Eg: CALibrate:MEAS:VOLTage 35,80,36.318,34.636,76.753,74.428)
Note: if Vm2 >= point2 or Vsps2 >= point2, if Vm2 > Vsps2, then
point2 = Vm2 + 0.1, otherwise point2 = Vsps2 + 0.1
(13) Send command CALibrate:SAVE CH1,VMEAS to save the calibration
data.
(14) Send command CALibrate:STATe CH1,0 to exit calibration mode.
(15) Voltage calibration completed.
The parameters corresponding to the different SPS models with the above
codes are shown in the following table
Mode
Rated
voltage(V)
Vset1
Vset2
Vset3
point1
point2
point3
SPS5041X
40
0.25
18
38
0
18
45
SPS5042X
SPS5043X
SPS5044X
SPS5045X
SPS5051X
50
0.3
23
46
0
23
70
SPS5081X
80
0.5
35
75
0
35
90
SPS5082X
SPS5083X
SPS5084X