46
set to default values (see Table 7.3.1).
The example below configures the DC voltage measurement, internally triggering the instrument to take
measurements, and reading the measurement results. The default range (automatic adjustment range) and
resolution (10 PLC) are used for measurement.
MEAS: VOLT: DC?
Typical response: +4.23450000E-03
The example below configures the instrument for 2-wire resistance measurements, triggers the instrument to take
measurements and reads the measurements. Select the 1 kΩ range.
MEAS: RES? 1000
Typical response: +3.2702
7.2.2.1 MEASure:VOLTage:{AC|DC}?
Description:
Configure the instrument for AC or DC voltage measurement and immediately trigger the
measurement to send the measurement to the user. All other measurement parameters and trigger parameters are
set to default values.
Syntax:
MEASure:VOLTage:{AC|DC}? [{<range>|AUTO|MIN|MAX|DEF}]
Parameters:
AC: <range>:{100m|1|10|100|750}. Default: AUTO
DC: <range>:{100m|1|10|100|1000}. Default: AUTO
Example:
MEAS:VOLT:DC? 100
Typical response: +8.531
7.2.2.2 MEASure:CURRent:{AC|DC}?
Description:
Configure the instrument for AC or DC current measurement and immediately trigger the
measurement to send the measurement to the user. All other measurement parameters and trigger parameters are
set to default values.
Syntax:
MEASure:CURRent:{AC|DC}? [{<range>|AUTO|MIN|MAX|DEF}]
Parameters:
<range>: {100u|1m|10m|100m|1|3|10}. Default: AUTO
Example:
MEAS:CURR:DC? 10
Typical response: +8.5301
7.2.2.3 MEASure:CONTinuity?
Description:
Configure the instrument to perform a continuity measurement and immediately trigger the
measurement to send the measurement to the user. All other measurement parameters and trigger parameters are
set to default values.
Syntax:
MEASure:CONTinuity?
Parameters:
(none)
Example:
MEAS:CONT?
Typical response: +8.5300