9.4.2.2. Read command
Refer to "15 Function setting lists" for commands for read setting values.
The following are commands for reading measured values.
Command
Description
1001
Read displayed value
1003
Read gross value
1005
Read net value
1007
Read tare value
1095
Read load cell input voltage (1nV/V)
The decimal point position is ignored when the setting value / measured value is read.
For example, the displayed value "123.4" is read as "1234".
Use the following procedure to read a setting value.
1. Set the command "0" in the "Command" area on PLC.
2. When AD-4411 accepts the command "0", it echoes back "0" in the "Response" area and waits for a
command from PLC.
3. Set a valid read command in the "Command" area on PLC.
4. If AD-4411 successfully reads the data, it echoes back the command in the "Response" area and updates
the read data in the "Response value" area. If it fails, the "Response" area is updated to 65535: Error.
The following is an example of reading FncF10 and setting value "35".
Command Value: N/A
Command: 0
AD-4411
PLC
To accept "0" command
Response Value: 0
Response: 0
To read FncF10
Command Value: 35
Command: 1319 (=reading FncF10)
Response Value: 35
Response: 1319 (=reading FncF10)
Command waiting state