Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012
7-31
Usage
statusInformation
=
bufferVar
.statuses[
N
]
statusInformation
The status value when reading
N
of the specified buffer was acquired
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
N
The reading number (
1
to
bufferVar
.n
)
Details
This read-only buffer recall attribute is like an array (a Lua table) of the status values for all the readings in the
buffer. The status values are floating-point numbers that encode the status value; see the following table for
values.
For dedicated reading buffers, all buffer attributes are saved to nonvolatile memory only when the reading buffer
is saved to nonvolatile memory.
Buffer status bits
Bit Name
Hex
value
Description
B0
FastADC
0x01
Fast ADC was used to make the reading
B1
Overtemp
0x02
Over temperature condition
B2
AutoRangeMeas
0x04
Measure range was autoranged
B3
AutoRangeSrc
0x08
Source range was autoranged
B4
4Wire
0x10
4-wire (remote) sense mode enabled
B5
Rel
0x20
Relative offset applied to reading
B6 Compliance
0x40 Source function was limited because the complementary
function would be over the compliance limit
B7 Filtered
0x80 Reading
was
filtered
Also see
(on page 7-23)
(on page 7-27)
(on page 7-32)
(on page 3-6)
bufferVar.timestampresolution
This attribute contains the timestamp's resolution.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
See
Details
1e-6 (1
μ
s)
Usage
resolution
=
bufferVar
.timestampresolution
resolution
Timestamp resolution in seconds
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
Summary of Contents for 2657A
Page 3: ......