Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012
7-353
tsplink.trigger[N].pulsewidth
This attribute sets the length of time that the trigger line is asserted for output triggers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
TSP-Link trigger N reset
Recall setup
Not saved
10e-6 (10
μ
s)
Usage
width
= tsplink.trigger[
N
].pulsewidth
tsplink.trigger[
N
].pulsewidth =
width
width
The pulse width (in seconds)
N
The trigger line (1 to 3)
Details
Setting the pulse width to 0 (seconds) asserts the trigger indefinitely.
Example
tsplink.trigger[3].pulsewidth = 20e-6
Sets pulse width for trigger line 3 to 20
μ
s.
Also see
(on page 7-353)
tsplink.trigger[N].release()
This function releases a latched trigger on the given TSP-Link trigger line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[
N
].release()
N
The trigger line (1 to 3)
Details
Releases a trigger that was asserted with an indefinite pulse width. It also releases a trigger that was latched in
response to receiving a synchronous mode trigger.
Example
tsplink.trigger[3].release()
Releases trigger line 3.
Also see
(on page 7-349)
(on page 7-356)
Summary of Contents for 2657A