Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012
7-329
Figure 120: SweepVLogMeasureI()
Example
SweepVLogMeasureI(smua, 100,
1000, 0.02, 5)
This function performs a five-point logarithmic voltage
sweep starting at 100 V and stopping at 1000 V. Current
is measured at every step (point) in the sweep after a
20 ms source settling period.
The following table contains log values and corresponding
source levels for the five-point logarithmic sweep:
MeasurePoint LogStepSize
LogStep SourceStepLevel
1
0.25
0.0
100.00 V
2
0.25
0.25
177.83 V
3
0.25
0.5
316.23 V
4
0.25
0.75
562.34 V
5
0.25
1.0
1000.0 V
Also see
(on page 5-19)
timer.measure.t()
This function measures the elapsed time since the timer was last reset.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
time
= timer.measure.t()
time
The elapsed time in seconds (1
μ
s resolution)
Example 1
timer.reset()
-- (intervening code)
time = timer.measure.t()
print(time)
Resets the timer and measures the time since the
reset.
Output:
1.01
The output will vary. The above output indicates that
timer.measure.t()
was executed 14.69077
seconds after
timer.reset()
.
startv
= Current measurement
stopv
stime
stime
stime