Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-17
Command Description
Default
parameter
SCPI
:IREFerence <b>
Enable or disable amps reference
:IREFerence?
Query state of amps reference
:DIGits <n>
Specify measurement resolution (4 to 7)
6
:AUTO <b>
Enable or disable auto resolution
:AUTO ONCE
Enable and then disable auto resolution
:AUTO?
Query auto resolution
:DIGits?
Query resolution
:AVERage
Path to control average filter
:TYPE <name>
Select filter type (
NONE, SCALar,
ADVanced
)
SCALar
:TYPE?
Query filter type
:TCONtrol
<name>
Select filter type: (
MOVing
or
REPeat
)
(Note 3)
:TCONtrol?
Query filter type
:COUNt <n>
Specify filter count (1 to 100)
10
:COUNt?
Query filter count
:ADVanced
Path to configure and control advanced
filter
:NTOLerance
<n>
Specify noise tolerance level (0 to 100 %) 1
:NTOLerance?
Query noise tolerance level
[:STATe] <b>
Enable or disable digital filter
OFF
[:STATe]?
Query state of digital filter
:RESistance
:MEDian
Path to control median filter
[:STATe] <b>
Enable or disable median filter
ON
[:STATe]?
Query state of median filter
:RANK <NRf>
Specify range (1 to 5)
1
:RANK?
Query rank
:DAMPing <b>
Enable or disable damping
OFF
:DAMPing?
Query state of damping
:VSControl <name>
Select ohms V-source mode (
MANual
or
AUTO
)
MANual
:VSControl?
Query ohms V-source mode
:MSELect <name>
Select ohms measurement type (
NORMal
or
RESistivity
)
NORMal
:MSELect?
Query ohms measurement type
:RESistivity
Path to configure resistivity:
:STHickness
<NRf>
Volume; specify sample thickness (0.0001
to 99.9999mm)
1mm
:STHickness?
Query sample thickness
:FSELect <name>
Specify test fixture (
M8009
or
USER
)
M8009
:FSELect?
Query test fixture
:M8009
Path to query Model 8009 test fixture