Programming with MR Protocol
248083227-1 Rev 8
6-13
Data Record
Commands
The Data Record Commands are used to transfer count data from the
particle counter to the computer and to control the instrument’s buffer
memory.
The syntax for the Data Record commands is a single capital letter with
no arguments.
Send Buffered Record(A)
The instrument sends the next record in its Rotating Buffer. The record
is erased from the buffer as it is sent.
Command:
A
Response:
A[Data stream
]
{Contents of next
buffered record}
-or-
A#
{Buffer empty}
Send Current Record(B)
The instrument sends the data record of the most recent sample period
in the Rotating Buffer. Data is not erased from the buffer.
Command:
B
Response:
B[Data Stream]
{Contents of most current record}
-or-
B#
{Buffer empty}
Clear Buffer(C)
Erases all records from the instrument’s Rotating Buffer.
Command:
C
{Buffer is cleared}