RLPC MODBUS Register Map v1.44
248083227-1 Rev 8
B-3
40020
ASCII string
Model Name char[10], char [11]
40021
ASCII string
Model Name char[12], char [13]
40022
ASCII string
Model Name char[14], char [15]
40023
unsigned integer
Value = Flow Rate. For example, 100d = 100.
40024
unsigned integer
Record Count. Total number of records stored in the counter.
40025
unsigned integer
Record Index. Zero based index to data in 30xxx register
series. Must be lower than the record count (register 40024).
Set this index to expose a counter’s record in the 30xxx
registers. Set to -1 to retrieve last record stored in the counter.
40026
unsigned integer
Location number.
Particle Counters: Specifies location of Particle Counter. Must
be 1 to 200 (maps to location names associated with registers
40200 - 40999).
Manifold Controller: Specifies Manifold position. Values 1-32
for the Universal Manifold and values 1-6 for the MiniManifold
Controller moves the arm to that position on the manifold.
Value 0 moves arm to Home position.
40027
signed integer
Real Time Clock (RTC) [high]. Updates instrument’s real-time
clock. Works in conjunction with 40028. Displays date and
time, in number of seconds since midnight, 1/1/1970. Can be
generated by ANSI C/C++ time() function.
40028
signed integer
Real Time Clock [low]
40029
unsigned integer
Initial Delay [high]. Works in conjunction with 40030.
Number of seconds to wait before starting the first sample. Max
value is 359,999, which equals 99h 59m 59s.
40030
unsigned integer
Initial Delay [low]
40031
unsigned integer
Hold Time [high]. Works in conjunction with 40032. Number
of seconds to wait between sample periods. Max value is
359,999, which equals 99h 59m 59s.
40032
unsigned integer
Hold Time [low]
40033
unsigned integer
Sample Time [high]. Works in conjunction with 40034.
Number of seconds to sample. Max value is 86,399, which
equals 23h 59m 59s.
40034
unsigned integer
Sample Time [low]
Table B-3 Sensor Settings Registers
Register Data
Type
Description
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......