Output #1 Submenus Options/Descriptions
PARAMETERS
(Flow Direction Mode)
OUTPUT
SOURCE
*FLOW RATE
DIRECTION
FORWARD ON
*REVERSE ON
Select whether the output is active when the flow is forward or reverse When
the absolute value of the flow rate is below the cutoff, the output will not be
active
PARAMETERS
(Alarm Mode)
ALARM
HIGH FLOW
LOW FLOW
OUT OF RANGE
*ERRORS ONLY
ALL
Select the flow condition or meter condition to trigger the
alarm and turn on the output
SET HIGH
Numeric entry Units and decimal
place based on FLOW RATE
selected Negative numbers
accepted Default is 1
Enter the value that the flow rate must be greater than in order
to trigger an alarm
SET HIGH
is only visible/settable when
ALARM
is set to
HIGH FLOW, OUT OF RANGE
or
ALL
SET LOW
Numeric entry Units and decimal
place based on FLOW RATE
selected Negative numbers
accepted Default is 0
Enter the value that the flow rate must be less than in order to
trigger an alarm
SET LOW
is only visible/settable when
ALARM
is set to
LOW FLOW, OUT OF RANGE
or
ALL
LATCHING
*DISABLED
ENABLED
When
ENABLED
, the output remains on after the alarm condition clears Resetting
alarm latch turns off the output
ANTI-
CHATTER
SET DELAY
Enter how long the alarm condition must occur before activating the output to
prevent nuisance trips Numeric entry Default is 100 ms
RELEASE
DELAY
Enter how long the alarm condition is cleared before resetting the output to
prevent the output from chattering The parameter is only valid if
LATCHING
is
DISABLED
Numeric entry Default is 100 ms
MIN ON-TIME
Numeric entry Default is 200 ms
PULL UP RESISTOR
INTERNAL
*EXTERNAL See
“Digital I/O Wiring” on page 15
Setup > Inputs/Output >Input
An asterisk (*) indicates the parameter default
Input Submenus
Options/Descriptions
MODE
DISABLED
Select the action to take when the input is active (based on the state)
*RESET FLOW TOTAL
UNLATCH ALARM
STATE
*ACTIVE ON HIGH
Select the voltage level to make the input active
ACTIVE ON LOW
Parameter Descriptions by Menu
Page 27
May 2021
TTM-UM-02537-EN-06