4-16
KFM2151
Remote Control
CHAN:VOLT:NSEL
Specifies the logical channel for voltage scanning. However, the logical channel
cannot be specified, if the specified channel is not a scanned source.
Command
CHANnel:VOLTage[:DC]:NSELect <chan>
CHANnel:VOLTage[:DC]:NSELect?
Parameter
<chan>
The default value is 1.
If the specified channel is not a voltage measurement source, an
SCPI error (-221, "Settings conflict") is returned.
The settings after sending *RST are indicated in Table 4-3.
Response
Returns the logical channel of the specified voltage scanning in the <NR1> form in
response to CHAN:VOLT:NSEL?.
CHAN:ROUT:TERM
Specify the logical channel for impedance scanning and set the terminal and the
scanned source (on/off). Auto scanning cannot be executed in remote mode.
Command
CHANnel:ROUTe:TERMinals
<chan>,<term+_NR1>,<term-_NR1>,{ON|OFF|1|0}
CHANnel:ROUTe:TERMinals? <chan>
Parameter
<term+_NR1>, <term-_NR1>
V
al
u
e:
1 to 32
The lower two digits of the specified logical channel
by default.
Parameter
{ON|OFF|1|0}
V
al
u
e:
ON(1)
Scan the specified channel (default)
OFF(0)
Not scan the specified channel
The settings after sending *RST are indicated in Table 4-3.
Response
Returns TERM+ and TERM- of the specified logical channel and the scanned
source (on/off) in the <NR1>,<NR1>,<NR1> form in response to
CHAN:ROUT:TERM? <chan>.
(Example)
If channel 1 of unit number 0 is term+7 and term-8 and is set as a scanned
source, the following is returned in response to CHAN:ROUT:TERM? 1.
7,8,1
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......