Bit Error Rate Tester | BERT 1005
Quantifi Photonics Ltd.
Version
2.04
83
Take the following steps to acquire eye scan measurement:
1.
Set the vertical resolution step size of the eye measurement to the maximum value:
:SENSe10:MEASure2:EDIagram:VERTical MAX
2.
Start the eye scan measurement:
:SENSe10:MEASure2:EDIagram:STARt
3.
Check the eye scan measurement status:
:CALCulate10:DATA2:EDIagram:STATus?
4.
Check the vertical resolution step size of the eye measurement:
:SENSe3:MEASure2:EDIagram:VERTical?
5.
Check the eye scan measurement data:
:CALCulate10:DATA2:EDIAGram?
6.
Stop the eye scan measurement:
:SENSe10:MEASure2:EDIagram:STOP
Take the following steps to configure and control pattern generation:
1.
Set the pattern type to PRBS9:
:SOURce10:PATTern2:TYPE PRBS9
2.
Set the pattern de-emphasis cursor to pre-cursor:
:SOURce10:PATTern2:DEEMphasis:CURSor PRE
3.
Set the pattern de-emphasis to 1.9 dB:
:SOURce10:PATTern2:DEEMphasis 1.9
4.
Start pattern generation:
:OUTPut10:DATA2:OUTPut 1
5.
Check the current SET state of the pattern generation output:
:OUTPut10:DATA2:OUTPut? SET
6.
Stop pattern generation:
:OUTPut10:DATA2:OUTPut 0